Start Chromatography system

Model Start is a compact, stand-alone preparative chromatography system designed for laboratory-scale protein purification, capable of efficiently performing fast, reliable one-step purification of a wide range of proteins. The system supports quick start methods, predefined templates and customized methods, meeting the automation requirements needed to provide high purity. It is easy to operate with an intuitive design, a visible flow path and a touchscreen interface, and offers diversified options for sample injection and automated fraction collection to further improve its performance.

Product Features

Compact, automated and easy-to-use protein purification.

l Equipped with an automatic injection valve to meet diverse sample loading and purification process requirements.

l Compact layout for fast, stable and reliable one-step protein purification.

l Diversified options for sample injection and automated fraction collection.

l Built-in quick start methods and professional templates for mainstream protein purification techniques.

l Intuitive touchscreen panel supporting real-time data monitoring and status observation.




  • Technical Details
  • URS
  • Download

Model

 Start

System Pump

Pump Type

10 mL peristaltic pump (single channel)

Flow rate

0.5-5 mL/min (working flow)
10 mL/min (wash flow)

Pump accuracy

±1.2%

Pressure range

0-10 bar

Viscosity range

0.6-5 cP

Mixer

Type

Static mixer

volume

0.4 mL

Valves

buffer valve

3-port solenoid switching valve,2 in 1 out

sample valve

3-port solenoid switching valve,2 in 1 out

outlet valve

3-port solenoid switching valve,1 in 2 out

automatic Injection valve

Rotary-type auto valve,6 modes

Pressure sensor

Post-pump pressure sensor
0-10 bar
±0.5 bar

Gradient formation

0.5-5ml/min
±5%

UV Detector

Wavelength range

280+260 nm, ±3 nm, dual wavelength

Absorbance range

-0.1 - +2 AU

Linear range

Within ±5%, up to 1.5 AU

Operating pressure

0-5 bar

Flow cell

2 mm path length (5 mm optional)
2 µL

Conductivity Detector

Detection range

0.001-999.99 mS/cm, ±2%

Flow cell

22 µL

Operating pressure

0-5 bar

Temperature detection range

0-99 °C

Temperature accuracy

±1.5 °C (within 4-45 °C)

pH Detector

Optional

Flow Restrictor

Optional

Fraction Collector
(Optional)

Compatible tube sizes

Supports 6 tube sizes

Collection volume range

0.5-50 mL

Accessories

workstation

software

Installation toolkit

Includes PEEK/PTFE pipe, pipe connectors, column clamp, back pressure valve, installation manual, common tools